メール
cjmarketing@163.com
-
電話番号
18018845013
-
アドレス
No. 528 Taiyun Road, Jiading District, Shanghai
Shanghai Changji Geological Instrument Co., Ltd
cjmarketing@163.com
18018845013
No. 528 Taiyun Road, Jiading District, Shanghai
Asphalt Bent Beam Rheometer (BBR)Suitable for determining the low-temperature bending creep stiffness of asphaltS(t)And creep slopemValue can be used to evaluate the low-temperature fracture characteristics of asphalt binders. This instrument is suitable for original asphalt, pressure aged asphalt, and aged asphalt after film oven (or rotary film oven).
1Asphalt Bent Beam Rheometer (BBR)Main technical features and functions
1. The instrument control unit adopts an industrial tablet computer, which displays deformation curves, pressure curves, and temperature curves in real time. The instrument clock software can be set and adjusted.
2. AdoptingPELTIETechnology and refrigeration combined mode, fast cooling speed,40It can be reduced from room temperature to room temperature in minutes-18℃, high temperature control accuracy.
3. Adopting linear differential displacement sensor(LVDT)Equipped with a standard calibration module, displacement parameters can be calibrated accurately and with high precision.
4. Adopting a high-precision, low-noise, and split refrigeration unit, and connecting it to the main unit with a hose to prevent its vibration from affecting the test results.
5. Adopting high-precision load sensors and equipped with standard pressure calibration devices, the pressure can be calibrated accurately and reliably.
6. The host liquid bath is equipped with a magnetic stirrer with adjustable speed, ensuring accurate and uniform temperature of the liquid bath.
7. The air source pressure, pre pressure, load pressure, and air bearing pressure can all be visually displayed in the operating software for easy viewing.
8. Design hasUSBCommunication port, capable of exporting instrument detection data.
9. It has functions such as instrument overload protection and fault warning. When an abnormality occurs, the buzzer will sound an alarm and the interface fault code will be displayed simultaneously.
10. The test time can be set for easy verification240sThe deformation situation after (default)240s).
11. The time required for calculating the results can be set by software (default)8s,15s,30s,60s,120s,240s).
12. Adopting two-stage high-precision air filters to provide high-quality compressed air for the air suspension system, ensuring detection accuracy.
IIMain technical indicators and parameters
1. Host sensor
Displacement sensor range:0~10)mm
Resolution of displacement sensor:5μm
Range of pressure sensor:0~4900)mN
Resolution of pressure sensor:2mN
Temperature sensor range:-50℃~Room temperature
Temperature sensor accuracy:±0.1℃
2. Host liquid bath
Temperature range of liquid bath:-40℃~Room temperature
Temperature control accuracy: ±0.1℃
Liquid bath volume:5L
Stirring speed:25turn/minute
3Other hosts
Trial mold size: 127mm x 12.7mm x 6.35mm
Top radius of sample support: 3mm ± 0.3mm
Loading axis radius: 6.3mm ± 1.3mm spherical contact point
Data interface: USB
Working power supply: AC220V/50Hz; 1kW
Environmental temperature: Room temperature 25 ℃ (air conditioning must be used if necessary)
Environmental humidity: ≤ 70% (relative humidity)
Size: 610mm × 610mm × 620mm
Weight: 80kg